MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2016-11-14 for SPECTRUM VENTRICLEAR? DRAINAGE CATHETER SET N/A N-VVDC-01-ABRM manufactured by Cook Inc.
[59786500]
(b)(4). The event is currently under investigation.
Patient Sequence No: 1, Text Type: N, H10
[59786501]
It was reported during removal the catheter broke and remains in the patients cranium. No additional information was available at the time of this report.
Patient Sequence No: 1, Text Type: D, B5
[74270511]
Investigation? Evaluation: a review of the complaint history, drawing, instructions for use (ifu), manufacturing instructions, quality control, and trends of the device was conducted during the investigation. The complaint device was not returned therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed. There is no evidence to suggest the product was not made to specifications. The lot number of the device is not known; accordingly a review of the device history record could not be conducted. Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined. We will continue to monitor for similar complaints. Per the quality engineering risk assessment no further action is required.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1820334-2016-01306 |
MDR Report Key | 6098450 |
Report Source | OTHER |
Date Received | 2016-11-14 |
Date of Report | 2017-04-24 |
Date Mfgr Received | 2017-03-13 |
Date Added to Maude | 2016-11-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. RITA HARDEN |
Manufacturer Street | 750 DANIELS WAY |
Manufacturer City | BLOOMINGTON IN 47404 |
Manufacturer Country | US |
Manufacturer Postal | 47404 |
Manufacturer Phone | 8123392235 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SPECTRUM VENTRICLEAR? DRAINAGE CATHETER SET |
Generic Name | NHC CATHETER, VENTRICULAR (CONTAINING ANTIBIOTIC 0R ANTIMICROBIAL AGENTS) |
Product Code | NHC |
Date Received | 2016-11-14 |
Model Number | N/A |
Catalog Number | N-VVDC-01-ABRM |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COOK INC |
Manufacturer Address | 750 DANIELS WAY BLOOMINGTON IN 47404 US 47404 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2016-11-14 |