INTELLISPACE CC & ANESTHESIA 866072

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,user facility report with the FDA on 2016-11-15 for INTELLISPACE CC & ANESTHESIA 866072 manufactured by Philips Medical Systems.

Event Text Entries

[59910434] A follow up report will be submitted once the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


[59910435] A medication order of heparin was modified by a doctor (end date entered) but the system did not generate a red flag. Clinician didn't see the end date for the order and the patient was administered an extra dose of heparin resulting in patient bleeding.
Patient Sequence No: 1, Text Type: D, B5


[71257100] Philips applications specialists have indicated that customer did not use the correct procedure to stop an order. Customer should have right click on order, click on stop order (interrompre in (b)(6)) then edit the order and change the end date, then a red flag appears when changes are saved. (philips toolkit guide for icca revision f page 161 - discontinued orders): workflow error -the issue was fully explained to he customer the customer did confirm this was a workflow error and not a philips product malfunction there was no product malfunction.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1218950-2016-07207
MDR Report Key6103181
Report SourceFOREIGN,USER FACILITY
Date Received2016-11-15
Date of Report2016-11-09
Date of Event2016-11-09
Date Mfgr Received2016-11-09
Date Added to Maude2016-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. BETTY HARRIS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal01810
Manufacturer Phone9786871501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTELLISPACE CC & ANESTHESIA
Generic NameINTELLISPACE CC & ANESTHESIA
Product CodeDXJ
Date Received2016-11-15
Model Number866072
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS
Manufacturer Address3000 MINUTEMAN ROAD ANDOVER MA 01810 US 01810


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-11-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.