CLEARSIGN? AMPLIFIER FOR LABSYSTEM? PRO EP RECORDING SYSTEM M0042002022R0 2002022

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-11-16 for CLEARSIGN? AMPLIFIER FOR LABSYSTEM? PRO EP RECORDING SYSTEM M0042002022R0 2002022 manufactured by Boston Scientific - Fremont (ce).

Event Text Entries

[60040471] (b)(4). Device evaluated by mfr: it is indicated that the device will not be returned for evaluation. A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted. If there is any further relevant information from that review, a supplemental medwatch will be filed. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[60040472] It was reported that the amplifier was exposed to high current. During preventative maintenance of a clearsign" amplifier for labsystem" pro ep recording system it was noted that the amplifier was exposed to too high current. The power supplied in the ep lab to the clearsign" amplifier was between was 240v-250v. Multiple errors in the system log were noted. The isolation transformer in the lab was modified to produce correct output (230 v). The clearsign" amplifier was exchanged. No patient involved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2016-10155
MDR Report Key6106841
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-11-16
Date of Report2016-10-18
Date of Event2016-10-18
Date Mfgr Received2016-10-17
Date Added to Maude2016-11-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. LINDA LEIMER
Manufacturer StreetONE SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634941700
Manufacturer G1BOSTON SCIENTIFIC - FREMONT (CE)
Manufacturer Street47215 LAKEVIEW BLVD NORTH DOCK
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal Code94538
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLEARSIGN? AMPLIFIER FOR LABSYSTEM? PRO EP RECORDING SYSTEM
Generic NameAMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL
Product CodeDRQ
Date Received2016-11-16
Model NumberM0042002022R0
Catalog Number2002022
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - FREMONT (CE)
Manufacturer Address47215 LAKEVIEW BLVD NORTH DOCK FREMONT CA 94538 US 94538


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.