ILED 4028210

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-11-18 for ILED 4028210 manufactured by Trumpf Medizin Systeme Gmbh + Co. Kg.

Event Text Entries

[60848278] It's known that an improper installation, maintenance or a collision can lead to the spring arm cover falling off. The device was evaluated and repaired by our local partner.
Patient Sequence No: 1, Text Type: N, H10


[60848279] The joint cover on an ondal spring arm, attached to the trumpf medical surgical light came off during a surgery and fell, landing outside of the operative field. No patient injury occurred.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681407-2016-00014
MDR Report Key6112459
Date Received2016-11-18
Date of Report2016-10-19
Date of Event2016-10-19
Date Mfgr Received2016-10-19
Date Added to Maude2016-11-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFFEN ULBRICH
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer CitySAALFELD, SAALE 07318
Manufacturer CountryGM
Manufacturer Postal07318
Manufacturer Phone6715864140
Manufacturer G1TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer City07318 SAALFELD,
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameILED
Generic NameILED5 ALC +
Product CodeFQP
Date Received2016-11-18
Model Number4028210
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer AddressCARL-ZEISS STRASSE 7-9 07318 SAALFELD, GM


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-18

© 2024 FDA.report