EASY GO VAC PM65HG

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-11-22 for EASY GO VAC PM65HG manufactured by Precison Medical, Inc..

Event Text Entries

[60660422]
Patient Sequence No: 1, Text Type: N, H10


[60660423] These aspirators behave as if the battery is low/discharge when turning the unit on. Biomed tech called tech support and also looked at the operator's manual with no indication of the error we were experiencing. After opening the unit, it was noticed that there was a blown fuse. After the fuse was replaced, the unit started to work normal again. Since then, we have identified at least 7 more devices with the same problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6120328
MDR Report Key6120328
Date Received2016-11-22
Date of Report2016-11-07
Date of Event2016-11-07
Report Date2016-11-07
Date Reported to FDA2016-11-07
Date Reported to Mfgr2016-11-07
Date Added to Maude2016-11-22
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEASY GO VAC
Generic NamePUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)
Product CodeBTA
Date Received2016-11-22
Model NumberPM65HG
Device Availability*
Device Age1 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPRECISON MEDICAL, INC.
Manufacturer Address300 HELD DRIVE NORTHAMPTON PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.