ENDOGATOR SINGLE USE AUXILIARY WATER JET CONNECTOR FOR PENTAX GI ENDOSCOPES 100242

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-11-22 for ENDOGATOR SINGLE USE AUXILIARY WATER JET CONNECTOR FOR PENTAX GI ENDOSCOPES 100242 manufactured by Medivators.

Event Text Entries

[60748487] The facility called to report that the endogator auxiliary water jet connector for pentax gi endoscopes was leaking because oring on the plastic connector that connects to the endoscope was too big. It was reported that the physician got sprayed with irrigation fluid when he activated the irrigation system. There is potential for cross contamination exposure. Medivators quality assurance team inspected the returned connectors from this lot and the issue could not be duplicated. The water that is sprayed from the connector would have been from the sterile water bottle when used in conjunction with the irrigation pump. There was no reported injury to the physician. This complaint will continue to be monitored within medivators complaint system.
Patient Sequence No: 1, Text Type: N, H10


[60748488] The case states that a single use auxiliary water jet connector for pentax endoscopes was leaking and sprayed the physician performing the procedure. There is potential for cross contamination.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651395-2016-00007
MDR Report Key6123023
Date Received2016-11-22
Date of Report2016-11-22
Date of Event2016-10-25
Date Mfgr Received2016-10-25
Device Manufacturer Date2016-09-03
Date Added to Maude2016-11-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMEGAN DICKEY
Manufacturer Street14605 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533327
Manufacturer G1MEDIVATORS
Manufacturer Street3150 POLLOK DR
Manufacturer CityCONROE TX 77303
Manufacturer CountryUS
Manufacturer Postal Code77303
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameENDOGATOR SINGLE USE AUXILIARY WATER JET CONNECTOR FOR PENTAX GI ENDOSCOPES
Generic NameAUXILIARY WATER JET CONNECTOR
Product CodeFEQ
Date Received2016-11-22
Returned To Mfg2016-11-08
Model Number100242
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDIVATORS
Manufacturer Address3150 POLLOK DR CONROE TX 77303 US 77303


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.