DIRECTCHECK ACT-LR ABNORMAL QUALITY CONTROL DCJLR-A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-11-23 for DIRECTCHECK ACT-LR ABNORMAL QUALITY CONTROL DCJLR-A manufactured by Accriva Diagnostics.

Event Text Entries

[60867731] (b)(4). Method: actual device not evaluated. Dhr review was not performed as the complaint is unrelated to product performance or packaging. No testing methods performed. Results: no results available since no evaluation performed. Conclusions: human factors issue. Training deficiency. Device not returned. Accriva diagnostics has requested all data required for form 3500a.
Patient Sequence No: 1, Text Type: N, H10


[60867732] Healthcare professional reported that an end user sustained an injury during reconstitution of a directcheck quality control. This control is packaged in a glass ampule inside a crushable plastic dropped vial containing diluent. The end-user was wearing gloves but it is unknown whether or not she utilized the protective sleeve provided with the product. The purpose of the sleeve is to safeguard end users against potential injury during reconstitution of the control. The end user sustained a small cut to an unspecified hand, which was caused by a glass shard protruding through the dropper vial. No other information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002721930-2016-00015
MDR Report Key6126521
Report SourceHEALTH PROFESSIONAL
Date Received2016-11-23
Date of Report2016-11-16
Date of Event2016-11-16
Date Mfgr Received2016-11-16
Device Manufacturer Date2016-03-17
Date Added to Maude2016-11-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJON MCDERMED
Manufacturer Street6260 SEQUENCE DRIVE
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582632490
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIRECTCHECK ACT-LR ABNORMAL QUALITY CONTROL
Generic NameDCJLR-A
Product CodeGGN
Date Received2016-11-23
Model NumberDCJLR-A
Catalog NumberDCJLR-A
Lot NumberC6DLA014
Device Expiration Date2017-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerACCRIVA DIAGNOSTICS
Manufacturer Address6260 SEQUENCE DRIVE SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.