MICROTARGETING ELECTRODE FC2001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2016-11-27 for MICROTARGETING ELECTRODE FC2001 manufactured by Fhc, Inc,.

Event Text Entries

[60907104] Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[60907105] Product used: fc2001, lot # 221236 - microtargeting electrode. During the use of the microtargeting electrode on (b)(6) 2016, the patient had a cerebral hemorrhage. Event was described as follows: after a lead was positioned in the right side, miniature potential in the left side was measured. Multi-technique method was conducted, and 2 tracks of center and postero were recorded. Waves could not be confirmed at depth of stn target (stn waves seemed to appear partly but were unstable). Sn waves seemed to be identified at plus 1. 5mm to 5 mm on the postero side. Tracks were changed to latte and antelo to measure again. Although impedance declined at 1200 ohms on latte side and around 700 ohms on antelo side, the procedure was continued. At re-measurement, even small typical background waves were identified. The patient experienced motor function and consciousness level in depression on the left side of the body. The procedure was aborted. Ct scan images identified cerebral hemorrhage. The patient was placed under monitoring the customer did not consider that the reported product caused the event issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002250546-2016-00003
MDR Report Key6128176
Report SourceDISTRIBUTOR,FOREIGN
Date Received2016-11-27
Date of Report2016-11-14
Date of Event2016-10-27
Date Mfgr Received2016-10-31
Date Added to Maude2016-11-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CRAIG PAGAN
Manufacturer Street1201 MAIN STREET
Manufacturer CityBOWDOIN ME 04287
Manufacturer CountryUS
Manufacturer Postal04287
Manufacturer Phone2076665651
Manufacturer G1FHC, INC,
Manufacturer Street1201 MAIN STREET
Manufacturer CityBOWDOIN ME 04287
Manufacturer CountryUS
Manufacturer Postal Code04287
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROTARGETING ELECTRODE
Generic NameDEPTH ELECTRODE
Product CodeGZL
Date Received2016-11-27
Model NumberFC2001
Lot Number221236
Device Expiration Date2019-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFHC, INC,
Manufacturer Address1201 MAIN STREET BOWDOIN ME 04287 US 04287


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2016-11-27

© 2026 FDA.report