VERSACELL SAMPLE MANAGEMENT SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-11-28 for VERSACELL SAMPLE MANAGEMENT SYSTEM manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[60995757] The customer contacted the siemens customer care center (ccc). Siemens is investigating this issue.
Patient Sequence No: 1, Text Type: N, H10


[60995758] The operator of a versacell sample management system reported that patient samples were found in the wrong locations on the system. The customer looked up patient sample (b)(6). Versacell indicated that sample was in position 488 b2, however, sample (b)(6) was located in that position. When sample (b)(6) was queried, versacell returned a different location of 469 a5. There are no reports of patient intervention or adverse health consequences due to the incorrect location of patient samples.
Patient Sequence No: 1, Text Type: D, B5


[63930762] The initial mdr 2247117-2016-00084 was filed on november 28th, 2016. Additional information (11/28/2016): a siemens customer service engineer (cse) was dispatched to the customer site. After evaluation of the instrument, the cse verified robot alignments, bar code reader, and pick and place. The issue reported could not be duplicated. Upon follow up, the customer stated they have not had any additional errors. The cause of incorrect location of patient samples on the instrument is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2247117-2016-00084
MDR Report Key6130232
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-11-28
Date of Report2016-12-16
Date of Event2016-10-25
Date Mfgr Received2016-11-28
Date Added to Maude2016-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARGARITA KARAN
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street62 FLANDERS BARTLEY RD
Manufacturer CityFLANDERS NJ 07836
Manufacturer CountryUS
Manufacturer Postal Code07836
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVERSACELL SAMPLE MANAGEMENT SYSTEM
Generic NameCLINICAL CHEMISTRY GENERAL PURPOSE LABORATORY EQUIPMENT
Product CodeCGJ
Date Received2016-11-28
Model NumberVERSACELL SAMPLE MANAGEMENT SYSTEM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address62 FLANDERS BARTLEY RD FLANDERS NJ 07836 US 07836


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.