[61086538] The case reported that patients are experiencing abdominal pain and pressure after procedures where the endostratus co2 insufflator was being used. (b)(4) sales representative is in close contact with this facility. This facility was performing regular colon screenings and it was reported that the extent of the pain and pressure the patients experienced was similar to that of procedures done using normal room temperature air. There were no lasting effects or hospitalization of these patients. The endostratus co2 insufflator unit was returned for additional inspection and functionality tests by (b)(4) depot service department. The unit performed according to specification. There is no details confirming that the facility co2 connection or processing error could be the cause. This complaint will continue to be maintained by (b)(4) complaint handling system.
Patient Sequence No: 1, Text Type: N, H10


[61086539] The facility reports that patients are experiencing post procedure abdominal pain and pressure after using co2 insufflator.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651395-2016-00008
MDR Report Key6133860
Date Received2016-11-29
Date of Report2016-11-28
Date of Event2016-11-01
Date Mfgr Received2016-11-01
Device Manufacturer Date2015-02-12
Date Added to Maude2016-11-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMEGAN DICKEY
Manufacturer Street14605 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533327
Manufacturer G1MEDIVATORS
Manufacturer Street3150 POLLOK DR
Manufacturer CityCONROE TX 77303
Manufacturer CountryUS
Manufacturer Postal Code77303
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameENDOSTRATUS CO2 INSUFFLATOR
Generic NameCO2 INSUFFLATOR FOR ENDOSCOPE
Product CodeFCX
Date Received2016-11-29
Returned To Mfg2016-11-07
Model NumberEGA-501
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDIVATORS
Manufacturer Address3150 POLLOK DR CONROE TX 77303 US 77303


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-29

© 2026 FDA.report