ACCUTREND CHOLESTEROL 11418262171

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign report with the FDA on 2016-11-30 for ACCUTREND CHOLESTEROL 11418262171 manufactured by Roche Diagnostics.

Event Text Entries

[61098653] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[61098654] The customer received a questionable cholesterol result from accutrend plus meter serial number rr0040140 when compared to the result from an accuchek performa meter. The result from the accuchek performa was 5. 5 mmol/l. The result from the accutrend plus was 2. 1 mmol/l. The customer thought the accuchek performa result was closer to her usual result. The customer has not been tested in a laboratory. The customer was not adversely affected. The suspect meter and strips were requested to be returned for investigation.
Patient Sequence No: 1, Text Type: D, B5


[63734890] Retention material from lot 204048-00 was tested with control material on a reference meter. The retention testing results showed no abnormalities and all results were within the tolerance ranges. Further investigation was possible as no customer material was received for investigation. A handling issue could not be excluded as a possible cause.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-01863
MDR Report Key6134478
Report SourceCONSUMER,FOREIGN
Date Received2016-11-30
Date of Report2016-12-21
Date of Event2016-11-16
Date Mfgr Received2016-11-16
Date Added to Maude2016-11-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCUTREND CHOLESTEROL
Generic NameCHOLESTEROL (TOTAL) TEST SYSTEM
Product CodeCHH
Date Received2016-11-30
Model NumberNA
Catalog Number11418262171
Lot Number204048-05
ID NumberNA
Device Expiration Date2016-12-31
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-30