TRIAGE CARDIAC TRIPLE MARKER PANEL 97000HSEU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-11-30 for TRIAGE CARDIAC TRIPLE MARKER PANEL 97000HSEU manufactured by Alere San Diego, Inc..

Event Text Entries

[61451266] Investigation conclusion: investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[61451267] Report received of discrepant results for troponin i at a patient clinic. Event occurred in (b)(6). Patient # 1 below had a positive result for troponin i, but the follow was a negative result. (b)(6). No reported adverse patient sequela. This discordant results only appear for troponin i. The other cardiac enzymes results were coherent with patient clinic. No additional information provided.
Patient Sequence No: 1, Text Type: D, B5


[63284406] Investigation/conclusion: customer's complaint was not replicated with in-house testing of retain lot w62034. No issues with tni recovery were observed. Manufacturing batch records for lot w62034 were reviewed and found that the lot met release specifications. Further investigation was not possible since the customer did not return any samples for testing. Unable to determine a root cause from the available information. Product deficiency was not established.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2016-00696
MDR Report Key6135699
Date Received2016-11-30
Date of Report2016-11-14
Date of Event2016-11-03
Date Mfgr Received2016-12-02
Date Added to Maude2016-11-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE CARDIAC TRIPLE MARKER PANEL
Generic NameCARDIAC MARKER TEST
Product CodeNBC
Date Received2016-11-30
Model Number97000HSEU
Lot NumberW62034RB
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-30

© 2026 FDA.report