VIDAS MEASLES IGG 30219

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-01 for VIDAS MEASLES IGG 30219 manufactured by Biomerieux, S.a..

Event Text Entries

[61564575] A customer in (b)(6) notified biom? Rieux of discrepant results associated with the vidas? Measles igg (reference 30219) related to false negative results involving a neqas sample (distribution 3974/specimen 3401). The customer reported the sample was tested on their primary testing platform (novatec), which identified the sample as positive. However, the customer indicated testing the sample later on vidas? And obtained a result of equivocal (0. 53, - <0. 5 would be negative). An internal biom? Rieux investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


[66261865] This report was initially submitted following notification that a customer in the (b)(6) notified biom? Rieux of discrepant results associated with the vidas? Measles igg (reference (b)(4)) related to false negative results involving a neqas sample ((b)(6)). The customer reported the sample was tested on their primary testing platform (novatec), which identified the sample as positive. The customer indicated that subsequent testing of the sample via vidas? Obtained a result of equivocal (0. 53, <0. 5 would be negative). An internal biom? Rieux investigation was conducted. The thresholds and interpreted results for vidas? Measles igg assay are the following: < 0. 50 negative. => 0. 50 to < 0. 70 equivocal. => 0. 70 positive. The external quality assessment "eqa msg 3401 from neqas distribution 3974" indicates that 88% of the participants who used vidas? Measles igg (msg) ref. 30219 obtained equivocal or negative results, though it was expected to be positive. Three (3) similar complaints were recorded for discrepant results using two (2) different vidas? Msg batches. - vidas msg lot 1004949820 (170430-0) gave an equivocal result (0. 68 tv) (pr# (b)(4)). - vidas msg lot 1004802610 (170302-0) gave an equivocal result (0. 53 tv) (pr# (b)(4)) and a negative result (0. 48 tv) (pr#(b)(4)). Analysis of the associated batch history records shows no anomaly during the manufacturing and control processes. The biom? Rieux quality product laboratory studied the control charts of six (6) internal samples (target : 0. 2 vt, 0. 17 vt, 3. 09 vt, 4. 81 vt, 1. 59 vt and 0. 31 vt) with eight (8) different batches of vidas? Msg including the two (2) customer batches: all results were within specifications. The quality product laboratory also tested the six (6) internal samples with vidas? Msg lots 170302-0 and 170430-0. The results were similar to those obtained during the release of the batches confirming that there is no evolution of the vidas? Msg lots 170302-0 and 170430-0. The quality product laboratory tested the specimen (b)(6) from neqas distribution 3974 with vidas? Msg lot 170302-0, lot 170430-0 and an additional batch 170622-0 (batch manufactured with different critical raw material) and obtained the following results: - vidas? Msg lot 170302-0 ; sample (b)(6): 0. 47 tv (584 rfv ) negative interpretation. Customers' results: 0. 53 tv negative interpretation and 0. 58 vt equivocal interpretation. The customer's results and results obtained at the quality product laboratory are very close from the equivocal lower zone (0. 5 tv) for this sample and are reproducible between themselves. It's written in the ifu that "equivocal samples should be repeated with a fresh specimen". - vidas? Msg lot 170430-0 ; sample (b)(6): 0. 66 tv (900 rfv) equivocal interpretation customers' results: 0. 68 tv equivocal interpretation the customer's results and results obtained at the quality product laboratory are reproducible between themselves. It's written in the ifu that "equivocal samples should be repeated with a fresh specimen". - vidas? Msg lot 170622-0 ; sample (b)(6): 0. 92 tv (1249 rfv) positive interpretation. External quality controls are manufactured samples and not natural samples from patients. Manufacturing process can affect sample matrix. No information was received from neqas regarding the manufacturing of this external quality control. Without this information, it's not possible to investigate further. Furthermore, internal samples close from the equivocal zone have been tested on these three (3) lots during the quality control of these lots and all gave the compliant interpretation, no interpretation change was observed from batch to batch. Based on the complaint trend analysis, on the results obtained on six (6) internal samples retested, on the results obtained on the specimen 3401 from neqas distribution 3974 which are close from the equivocal lower zone or within the equivocal zone and on the fact that no information was received from neqas regarding the manufacturing of this external quality control, the investigation concludes that performance of vidas? Msg lots 170302-0 and 170430-0 are within specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00494
MDR Report Key6139152
Date Received2016-12-01
Date Mfgr Received2016-11-09
Date Added to Maude2016-12-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, S.A.
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS MEASLES IGG
Generic NameVIDAS MEASLES IGG
Product CodeLIB
Date Received2016-12-01
Catalog Number30219
Lot Number1004802610
Device Expiration Date2017-03-02
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, S.A.
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.