SYNGO LAB DATA MANAGER APPLICATION SOFTWARE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-12-02 for SYNGO LAB DATA MANAGER APPLICATION SOFTWARE manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[61400814] The customer contacted the siemens customer care center (ccc). Ccc connected remotely and checked a patient sample. Ccc discovered that data was not purged after 30 days as configured. A siemens headquarter support center (hsc) specialist connected remotely and performed a data base clean up, purge and defragmentation. Hsc checked the database, and no issues were noted. The cause of the database not purging as configured is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[61400815] The (b)(4) lab data manager was not purging data as configured, causing results to stay in the system past the time period configured. The customer reuses barcodes every 90 days, and discovered that old results were being confused with new orders and the same barcodes were associated with different patients. There were no incorrect results or data reported out by the customer. There are no known reports of patient intervention or adverse health consequences due to the (b)(4) not purging data as configured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2016-00484
MDR Report Key6142011
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-12-02
Date of Report2016-12-02
Date of Event2016-11-03
Date Mfgr Received2016-11-03
Date Added to Maude2016-12-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARGARITA KARAN
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYNGO LAB DATA MANAGER APPLICATION SOFTWARE
Generic NameSYNGO LAB DATA MANAGER APPLICATION SOFTWARE
Product CodeNSX
Date Received2016-12-02
Model NumberSYNGO LAB DATA MANAGER APPLICATION SOFTWARE
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-02