OSSEOFLEX SN+ OF-0021

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-05 for OSSEOFLEX SN+ OF-0021 manufactured by Osseon Llc.

Event Text Entries

[61492566] Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[61492567] During the vertebral augmentation procedure, the device was left in the patient for approximately 15 minutes after the cement was injected. The delay caused the cement to harden and a portion of the distal needle broke off (less than 40mm) and remained in the patient as the needle could not be dislodged. No image was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007111981-2016-00002
MDR Report Key6146522
Date Received2016-12-05
Date of Report2016-12-05
Date of Event2016-11-11
Date Mfgr Received2016-11-11
Device Manufacturer Date2015-09-24
Date Added to Maude2016-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RON CLOUGH
Manufacturer Street2301 CIRCADIAN WAY SUITE 300
Manufacturer CitySANTA ROSA CA 95407
Manufacturer CountryUS
Manufacturer Postal95407
Manufacturer Phone7076365940
Manufacturer G1OSSEON LLC
Manufacturer Street2301 CIRCADIAN WAY SUITE 300
Manufacturer CitySANTA ROSA CA 95407
Manufacturer CountryUS
Manufacturer Postal Code95407
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOSSEOFLEX SN+
Generic NameSTEERABLE VERTEBRAL NEEDLE
Product CodeOAR
Date Received2016-12-05
Model NumberOF-0021
Lot Number081015-01
Device Expiration Date2017-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOSSEON LLC
Manufacturer Address2301 CIRCADIAN WAY SUITE 300 SANTA ROSA CA 95407 US 95407


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2016-12-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.