NEOTREND-L SENSOR N7004L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2004-07-23 for NEOTREND-L SENSOR N7004L manufactured by Diametrics Medical, Ltd..

Event Text Entries

[417819] Upon successful calibration, a neotrend-l sensor was attached to the vac and sensor insertion was attempted. The sensor would not penetrate past the vac hub. After many attempts, blood was seen backing up the vac and then into the sensor and out of the advancement lock. After close examination of the vac, it was noted that the vac was an argyle 3. 5 fr and not a dml 3. 7 fr. No intervention was required due to this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612233-2004-00015
MDR Report Key614775
Report Source06
Date Received2004-07-23
Date of Report2004-06-28
Date of Event2004-06-26
Date Mfgr Received2004-06-28
Date Added to Maude2005-06-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMISS J. WALTERS
Manufacturer StreetGEORGE STREET, HIGH WYCOMBE
Manufacturer CityBUCKS HP112XJ
Manufacturer CountryUK
Manufacturer PostalHP11 2XJ
Manufacturer Phone494471671
Manufacturer G1DIAMETRICS MEDICAL LIMITED
Manufacturer StreetSHORT STREET, HIGH WYCOMBE
Manufacturer CityBUCKS HP112QH
Manufacturer CountryUK
Manufacturer Postal CodeHP11 2QH
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNEOTREND-L SENSOR
Generic NameBLOOD GAS MONITORING SYSTEM
Product CodeCCE
Date Received2004-07-23
Model NumberN7004L
Catalog NumberN7004L
Lot NumberUNK
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key604532
ManufacturerDIAMETRICS MEDICAL, LTD.
Manufacturer AddressGEORGE ST. HIGH WYCOMBE, BUCKS. UK
Baseline Brand NameNEOTREND-L MULTIPARAMETER SENSOR
Baseline Generic NameNEOTREND
Baseline Model NoN7004L
Baseline Catalog NoN7004L
Baseline IDNA
Baseline Shelf Life ContainedY
Baseline Shelf Life [Months]18
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK972314
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2004-07-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.