MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-12-07 for ADVIA CENTAUR XP manufactured by Siemens Healthcare Diagnostics Inc..
[61734822]
The customer contacted the siemens customer care center (ccc). Siemens is investigating the issue.
Patient Sequence No: 1, Text Type: N, H10
[61734823]
A discordant, falsely depressed cancer antigen 125 (ca-125) result was obtained on one patient sample on an advia centaur xp instrument. The discordant result was reported to the physician(s), who questioned it. The sample was repeated on the same instrument, resulting higher. The corrected result was reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely depressed ca-125 result.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2432235-2016-00750 |
MDR Report Key | 6153238 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2016-12-07 |
Date of Report | 2016-12-07 |
Date of Event | 2016-11-02 |
Date Mfgr Received | 2016-11-14 |
Device Manufacturer Date | 2015-10-05 |
Date Added to Maude | 2016-12-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MARGARITA KARAN |
Manufacturer Street | 511 BENEDICT AVE |
Manufacturer City | TARRYTOWN NY 10591 |
Manufacturer Country | US |
Manufacturer Postal | 10591 |
Manufacturer Phone | 9145243105 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS |
Manufacturer Street | MANUFACTURING LIMITED REGISTRATION NUMBER: 8020888 |
Manufacturer City | CHAPEL LANE, SWORDS, CO., DUBLIN |
Manufacturer Country | EI |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ADVIA CENTAUR XP |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | LTK |
Date Received | 2016-12-07 |
Model Number | ADVIA CENTAUR XP |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 |
Brand Name | ADVIA CENTAUR XP |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2016-12-07 |
Model Number | ADVIA CENTAUR XP |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-12-07 |