CAUTERY HIGH TEMP FINE TIP AA01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2016-12-09 for CAUTERY HIGH TEMP FINE TIP AA01 manufactured by Bovie Medical Corporation.

Event Text Entries

[61863870] Distributer received a box of 9 of 10 cauteries with open pouch. Cauteries model aa01 sterile lot 0816j lot # 2016-08-06. The returned product was inspected and it was confirmed that the pouched units were unsealed. No additional complaints have been received for this catalog number aa01 manufacturing lot number 2016-08-06, sterile lot 0816j. Review of the batch record showed no non-conformance noted during the pouching process, packaging or qc inspection processes. Employees were made aware of the event and the sealing and packaging employees were instructed to assure adequate handling and separation of unsealed pouches.
Patient Sequence No: 1, Text Type: N, H10


[61863871] Customer received a box with 9 of 10 unsealed pouches.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007593903-2016-00012
MDR Report Key6157513
Report SourceDISTRIBUTOR
Date Received2016-12-09
Date of Report2016-12-09
Date of Event2016-11-11
Date Mfgr Received2016-11-11
Device Manufacturer Date2016-08-25
Date Added to Maude2016-12-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJULIE MCDONALD
Manufacturer Street5115 ULMERTON ROAD
Manufacturer CityCLEARWATER FL 337604004
Manufacturer CountryUS
Manufacturer Postal337604004
Manufacturer Phone7278038554
Manufacturer G1BOVIE MEDICAL CORPORATION
Manufacturer Street5115 ULMERTON ROAD
Manufacturer CityCLEARWATER FL 337604004
Manufacturer CountryUS
Manufacturer Postal Code337604004
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAUTERY HIGH TEMP FINE TIP
Generic NameAA01
Product CodeHQP
Date Received2016-12-09
Returned To Mfg2016-11-17
Catalog NumberAA01
Lot Number0816J
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOVIE MEDICAL CORPORATION
Manufacturer Address5115 ULMERTON ROAD CLEARWATER FL 337604004 US 337604004


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.