Event Text Entries

[62376589] Investigation conclusion: customer's complaint was not replicated with in-house testing of retain lot w62131. No issues with d-dimer recovery were observed. Manufacturing batch records for lot w62131 were reviewed and found that the lot met release specifications. Further investigation was not possible since the customer did not return any samples for testing. Unable to determine a root cause from the available information. Product deficiency was not established.
Patient Sequence No: 1, Text Type: N, H10


[62376590] Event occurred in (b)(6). The customer reported receiving a false negative result with triage d-dimer panel from lot w62131rba. The customer received a negative result of <100. The customer tested with edta whole blood. The other gp of the patient sent a sample for confirmatory testing to a lab on the same day and they received a positive result of 694. The patient's specimen was sent to the lab in a citrate tube. There were no reported adverse events related to the triage result. No additional information provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2016-00701
MDR Report Key6159947
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-12-09
Date of Report2016-11-30
Date Mfgr Received2016-11-30
Date Added to Maude2016-12-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE D-DIMER PANEL
Generic NameCARDIAC MARKER TEST
Product CodeNBC
Date Received2016-12-09
Model Number98100EU
Lot NumberW62131RBA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-09

© 2026 FDA.report