VITROS CHEMISTRY PRODUCTS OP REAGENT 6801997

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-13 for VITROS CHEMISTRY PRODUCTS OP REAGENT 6801997 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[62677558] The investigation determined that a discordant, negative vitros op-lo result was obtained from a single patient sample run on the vitros 5600 integrated system. The intended use section of the vitros op ifu states that the vitros chemistry product op assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with the vitros op assay. Gas chromatography/mass spectrometry (gcms) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result. The likely cause is a known limitation of the vitros op reagent related to low cross reactivity with oxymorphone and oxycodone. Although there is no indication that the instrument was not performing as intended, an instrument issue cannot be completely ruled out as a contributing factor to the event. There is no information to suggest that the vitros op reagent malfunctioned.
Patient Sequence No: 1, Text Type: N, H10


[62677559] The customer obtained a discordant, negative vitros op-lo result from a single patient sample run on the vitros 5600 integrated system, when compared to a positive opiate result obtained using the gc/ms method. Vitros patient result: 279 ng/ml (positive cutoff >or = 300 ng/ml) vs. >positive (2500 ng/ml oxycodone and 620 ng/ml oxymorphone). Biased results of the magnitude and direction observed may lead to inappropriate physician action. The discordant vitros op-lo result was reported out of the laboratory. However, there was no report of treatment started, stopped, or altered based on the false negative vitros op-lo result and there was no report of patient harm as a result of this event. This report corresponds to ortho clinical diagnostics inc. Complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319808-2016-00042
MDR Report Key6169007
Date Received2016-12-13
Date of Report2016-12-13
Date of Event2016-11-17
Date Mfgr Received2016-11-28
Device Manufacturer Date2016-03-10
Date Added to Maude2016-12-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street1000 LEE ROAD
Manufacturer CityROCHESTER NY 14606
Manufacturer CountryUS
Manufacturer Postal Code14606
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS OP REAGENT
Generic NameIN VITRO DIAGNOSTIC
Product CodeDJG
Date Received2016-12-13
Catalog Number6801997
Lot Number1527-09-5274
Device Expiration Date2017-09-10
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.