CARBOFLO VASCULAR GRAFTS F5006TWSC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-12-14 for CARBOFLO VASCULAR GRAFTS F5006TWSC manufactured by Bard Peripheral Vascular, Inc..

Event Text Entries

[62382787] No device, no medical records, or no medical images were provided to the manufacturer. The lot number for the device was provided. The device history records are currently under review. The investigation of the reported event is currently underway. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
Patient Sequence No: 1, Text Type: N, H10


[62382788] It was reported that during an intraoperative procedure, an alleged tear in the vascular graft was identified upon removal of the beading. The health care provider removed the torn segment and completed the procedure with the vascular graft. There was no reported patient injury.
Patient Sequence No: 1, Text Type: D, B5


[66911400] Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing. This lot met all release criteria. There was nothing found to indicate there was a manufacturing related cause for this event. This is the only event reported to date for this lot number and failure mode. Visual inspection: a segment of the graft was returned with the product packaging and label. The graft segment had two blue lines running longitudinally and had carbon lining, which identified the product as a bard graft. The segment measured 4. 1cm in length at its longest point. The beading was missing along entire returned length. It appeared that one end of the returned graft was cut. The other end appeared to be torn. There was a tear along the beading track of the returned segment. Medical records review: medical records were not provided for review. Image/photo review: images/photos were not provided for review. Conclusion: the investigation was confirmed for torn material, as the returned segment had a tear along the beading track. It was unknown if procedural issues contributed to the reported event. Based on the available information, the definitive root cause was unknown. Labeling review: the current ifu states: precautions: when removing the external spiral support (beading) of the graft, the beading must be removed slowly and at a 90? Angle to the graft. Rapid unwinding and /or removal at less than a 90? Angle may result in graft damage. Do not use surgical blades or sharp, pointed instruments to remove the beading as this may damage the graft wall. If damage occurs, that segment of the graft should not be used. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
Patient Sequence No: 1, Text Type: N, H10


[66911401] It was reported that during an intraoperative procedure, an alleged tear in the vascular graft was identified upon removal of the beading. The health care provider removed the torn segment and completed the procedure with the vascular graft. There was no reported patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020394-2016-01173
MDR Report Key6173561
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-12-14
Date of Report2016-11-18
Date of Event2016-11-17
Date Mfgr Received2017-01-27
Device Manufacturer Date2015-12-13
Date Added to Maude2016-12-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJUDITH LUDWIG
Manufacturer Street1625 W 3RD ST.
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal85281
Manufacturer Phone4803032689
Manufacturer G1BARD PERIPHERAL VASCULAR, INC.
Manufacturer Street1625 W 3RD ST.
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal Code85281
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARBOFLO VASCULAR GRAFTS
Generic NameEPTFE VASCULAR GRAFT
Product CodeDYF
Date Received2016-12-14
Returned To Mfg2016-12-13
Catalog NumberF5006TWSC
Lot NumberVTZL0360
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBARD PERIPHERAL VASCULAR, INC.
Manufacturer Address1625 W 3RD ST. TEMPE AZ 85281 US 85281


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.