AMS 700 72402960

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-14 for AMS 700 72402960 manufactured by American Medical Systems, Llc.

Event Text Entries

[62366248]
Patient Sequence No: 1, Text Type: N, H10


[62366249] Patient was taken to surgery for replacement of a malfunctioning inflatable penile prosthesis that was placed at this hospital 10 years ago. Patient reported fluid loss pre-operatively. Surgeon noted that device had worked well until recently when patient noticed lost fluid. During the replacement surgery, surgeon noted that due to patient's obesity, the tubing tore and the reservoir was left in situ. There was space for the fluid reservoir and placed a low profile 100 ml inhibizone covered reservoir in this cavity. An 18 cm pre-connected cylinders with 5 cm rear-tip extenders were placed in each corporal body. There were no post-op complications and the patient was discharged to home the day after surgery. Ams worksheet lists the penile prosthesis as ams 700 tactile pump. The previous surgery lists the following implanted devices: ams 700(tm) reservoir, iz ref 72402960, lot # 433758004; volume 100 ml. Ams 700(tm) accessory kit, ref 72401850, lot # 441323006. Ams 700 cx(tm) preconnect, iz, ref 72403965, lot # 438038007; 18 cm x 12 mm. Manufacturer response for inflatable penile prosthesis, ams 700 tm (per site reporter): manufacturer will evaluate the device when it is returned.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6173669
MDR Report Key6173669
Date Received2016-12-14
Date of Report2016-10-19
Date of Event2016-10-11
Report Date2016-10-19
Date Reported to FDA2016-10-19
Date Reported to Mfgr2016-10-19
Date Added to Maude2016-12-14
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS 700
Generic NamePROSTHESIS, PENIS, INFLATABLE
Product CodeJCW
Date Received2016-12-14
Model NumberAMS 700
Catalog Number72402960
Lot Number433758004
Device AvailabilityY
Device Age10 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAMERICAN MEDICAL SYSTEMS, LLC
Manufacturer Address10700 BREN ROAD WEST MINNETONKA MN 55343 US 55343