AMIS LEG POSITIONER 01.15.10.0190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-12-16 for AMIS LEG POSITIONER 01.15.10.0190 manufactured by Medacta International Sa.

Event Text Entries

[62601269] Additional information received from the initial reporter on (b)(6) 2016 and includes: the nurse was? Expert? And used to manipulate the leg positioner but first the rotation was uneasy and then probably happened the fracture. However they finally noticed the fracture at the end of the intervention during the reduction. The fracture was fixed during the same intervention. On (b)(6) 2016 the medical affairs director performed a clinical evaluation and commented as follows: the succession of events is not very clear: apparently, a tibial fracture was generated during the tha operation in a very osteoporotic patient (according to report), possibly during the external rotation of the limb. This operation is done manually by an or help, normally. It is not clear how and why medacta devices would be responsible for this event. No final conclusion can be drawn with the information currently available. Batch reviews performed on (b)(6) 2016. Amis mobile leg positioner 2. 0, code 01. 15. 10. 0190, lot. 1411958 one item manufactured and released on (b)(6) 2014. No anomalies found related to the problem. Amis shoe, code 01. 15. 10. 0315 lot 1551837: the 30 items manufactured and released on (b)(6) 2015. No anomalies found related to the problem. To date, no similar events have been reported on items of the same lot.
Patient Sequence No: 1, Text Type: N, H10


[62601270] At the end of the hip surgery it was noticed that the tibia around malleolus of the patient was fractured. The nurses in the or first had a hard maneuver with external rotation.
Patient Sequence No: 1, Text Type: D, B5


[70218133] On 13 march 2017 the r&d project manager performed a visual inspection of the retrieved item and commented as follows: following visual and functional checks of the item, no improprierties were found.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005180920-2016-00669
MDR Report Key6180744
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-12-16
Date of Report2017-03-17
Date of Event2016-11-18
Date Mfgr Received2016-11-18
Device Manufacturer Date2014-11-27
Date Added to Maude2016-12-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFANO BAJ
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, 6874
Manufacturer CountrySZ
Manufacturer Postal6874
Manufacturer Phone91 6966060
Manufacturer G1MEDACTA INTERNATIONAL SA
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, 6874
Manufacturer CountrySZ
Manufacturer Postal Code6874
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMIS LEG POSITIONER
Generic NameTABLE FOR SURGERY
Product CodeFWX
Date Received2016-12-16
Returned To Mfg2017-02-24
Catalog Number01.15.10.0190
Lot Number1411958
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDACTA INTERNATIONAL SA
Manufacturer AddressSTRADA REGINA CASTEL SAN PIETRO, 6874 SZ 6874


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-12-16