WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
D
Patient 1
IN THIS EVENT IT WAS REPORTED THAT A PAIR OF PALODENT PLUS FORCEPS BROKE AT THE TIP; NO INJURY RESULTED.
N
Patient 1
RIGHT TIP HAS BROKEN OFF. FRACTURE SURFACE SHOWS SIGNS OF CORROSION WELL ACROSS THE SURFACE WHICH INDICATES THERE WAS LIKELY A CRACK IN THE BENT AREA FOR SOME TIME BEFORE THE BREAKAGE. THE LEFT TIP IS STILL INTACT BUT IT ALSO HAS A CRACK IN THE BEND WHICH WOULD ALSO HAVE FAILED SOON. INCORRECT MANUFACTURE OF THE TIP AREA IS THE MOST LIKELY CAUSE OF THE INITIAL CRACKING WHICH ULTIMATELY LEADS TO THE BREAKAGE. A SCAR HAS BEEN ISSUED TO THE ORIGINAL MANUFACTURER TO CORRECT THIS ISSUE.