SPS-1 STATIC PRESERVATION SOLUTION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-19 for SPS-1 STATIC PRESERVATION SOLUTION manufactured by Organ Recovery Systems, Inc..

Event Text Entries

[62696545]
Patient Sequence No: 1, Text Type: N, H10


[62696546] A liver transplant recipient anesthetized on table. When the sterile bag containing the liver was opened, the transplant surgeon immediately noted a foul odor (later became even more potent) and the case was ceased until the source of the odor could be determined. Case subsequently cancelled as cause of odor determined to be coming from at least two bags of organ preservation fluid. Samples of fluid were taken and sent to microbiology; results came back positive for a number of organisms. Preliminary results showed the following: sample [1] is original sample, directly out of bag perfusing one of the kidneys. Growth is abundant. Preliminary identifications so far are for enterococcus casseliflavus and pantoea agglomerans. Sample [2] is "sps straight out of bag into jar", and it looks clean by gram stain. Sample [3] is "sps from jar w/lt kidney" and has a gram stain similar to the original one from yesterday (mixed organisms, abundant). Sample [4] is "sps lot pbr 0060 392". Gram-positive cocci that are similar to those previously; for whatever reason they're more sparse and the gram negative organisms are not here but may appear later in culture. Sample [5] is "sps lot pbr 0074 330". This appeared clean by gram stain at the time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6184847
MDR Report Key6184847
Date Received2016-12-19
Date of Report2016-12-13
Date of Event2016-12-12
Report Date2016-12-13
Date Reported to FDA2016-12-13
Date Reported to Mfgr2016-12-13
Date Added to Maude2016-12-19
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPS-1 STATIC PRESERVATION SOLUTION
Generic NameSYSTEM, PERFUSION, KIDNEY
Product CodeKDL
Date Received2016-12-19
Lot NumberPBR0074330
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerORGAN RECOVERY SYSTEMS, INC.
Manufacturer AddressONE PIERCE PLACE SUITE 475W ITASCA IL 60143 US 60143


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.