ENVISION BLOWER KIT P136026

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-12-19 for ENVISION BLOWER KIT P136026 manufactured by Hill-rom Batesville.

Event Text Entries

[62746687] The pressure wound was treated with dakins solution and abdominal pads. Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface. Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy. Position changes are key to pressure sore prevention and treatment. These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas. The blower connection was not properly attached causing the blower to not function as designed. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this blower in august 2016. It is unknown if the facility performed any other preventative maintenance on this blower. The technician properly connected the blower to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[62746688] Hill-rom received a report from the account stating the patient developed a pressure ulcer on his sacrum that progressed to stage 4. The blower was located at the account. This report was filed in our complaint handling system as complaint # (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2016-00532
MDR Report Key6186387
Report SourceUSER FACILITY
Date Received2016-12-19
Date of Report2016-11-21
Date of Event2016-11-21
Date Mfgr Received2016-11-21
Date Added to Maude2016-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLORI DAULTON
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENVISION BLOWER KIT
Generic NameBED FLOTATION THERAPY POWERED
Product CodeIOQ
Date Received2016-12-19
Model NumberP136026
OperatorOTHER CAREGIVERS
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-12-19