NEOTREND-L SENSOR N7004L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2004-08-06 for NEOTREND-L SENSOR N7004L manufactured by Diametrics Medical, Ltd..

Event Text Entries

[426814] The neotrend-l sensor was placed and ran for six days with good correlation. Then, due to a thermocouple failure, plus the decision to withdraw treatment from this terminal patient, the sensor was removed. The vac was left in place. Upon retraction of the sensor it was noted that 3 cm of the pco2 sensor was left outside the y-piece with the sensor fully retracted. The tip of the neotrend-l sensor had become detached and was missing. The nurse claims that the uac was not clamped during sensor retraction. The sensor was returned to the mfr for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612233-2004-00017
MDR Report Key618840
Report Source06
Date Received2004-08-06
Date of Report2004-07-09
Date of Event2004-07-08
Date Mfgr Received2004-07-09
Device Manufacturer Date2003-10-01
Date Added to Maude2005-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMISS J. WALTERS
Manufacturer StreetGEORGE STREET, HIGH WYCOMBE
Manufacturer CityBUCKS HP112XJ
Manufacturer CountryUK
Manufacturer PostalHP11 2XJ
Manufacturer Phone494471671
Manufacturer G1DIAMETRICS MEDICAL LIMITED
Manufacturer StreetGEORGE STREET, HIGH WYCOMBE
Manufacturer CityBUCKS HP112XJ
Manufacturer CountryUK
Manufacturer Postal CodeHP11 2XJ
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNEOTREND-L SENSOR
Generic NameBLOOD GAS MONITORING SYSTEM
Product CodeCCE
Date Received2004-08-06
Returned To Mfg2004-07-12
Model NumberN7004L
Catalog NumberN7004L
Lot Number907
ID Number*
Device Expiration Date2005-05-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeNA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key608520
ManufacturerDIAMETRICS MEDICAL, LTD.
Manufacturer AddressGEORGE ST. HIGH WYCOMBE, BUCKS UK HP11 2XJ
Baseline Brand NameNEOTREND-L MULTIPARAMETER SENSOR
Baseline Generic NameNEOTREND
Baseline Model NoN7004L
Baseline Catalog NoN7004L
Baseline IDNA
Baseline Shelf Life ContainedY
Baseline Shelf Life [Months]18
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK972314
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2004-08-06

© 2026 FDA.report