COHERENCE DOSIMETRIST + ONCOLOIGST WORKSPORE 5856641 7333680 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2004-08-03 for COHERENCE DOSIMETRIST + ONCOLOIGST WORKSPORE 5856641 7333680 NA manufactured by Siemens Medical Solutions Usa, Inc..

Event Text Entries

[426931] An event did not occur, however an intermittent software anomaly has been discovered whereby imported data may be missing after performing a dicom export. The missing data may cause incorrect treatment planning for radiation therapy on third party vendor software.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2910081-2004-00001
MDR Report Key618860
Report Source07
Date Received2004-08-03
Date Mfgr Received2004-06-10
Device Manufacturer Date2002-11-01
Date Added to Maude2005-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactKEN NEHMER
Manufacturer Street4040 NELSON AVE
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal94520
Manufacturer Phone9252468200
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction Number2910081-08/04/04-005C
Event Type3
Type of Report3

Device Details

Brand NameCOHERENCE DOSIMETRIST + ONCOLOIGST WORKSPORE
Generic NameSW OPTION FOR MEDICAL CHARGED-PARTICLE RADIATION THERAPY
Product CodeHJW
Date Received2004-08-03
Model Number5856641 7333680
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key608540
ManufacturerSIEMENS MEDICAL SOLUTIONS USA, INC.
Manufacturer Address* CONCORD CA * US
Baseline Brand NameCOHERENCE DOSIMETRIST + ONCOLOIGST WORKSPORE
Baseline Generic NameSW OPTION FOR MEDICAL CHARGED-PARTICLE RADIATION THERAPY
Baseline Model No5856641 7333680
Baseline Catalog NoNA
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2004-08-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.