[63312870] The reamer was returned for evaluation. As returned, the awl is bent and has fractured at the base of the handle. Review of receiving inspection report for the device indicates the devices were manufactured to specifications. Inspection documentation shows all devices that were inspected met specifications. Wear & tear is seen that indicates successful use during a potential field age of approximately 10 years 10 months. Dimensional measurements are conforming to print specifications. The instrument was manufactured on september 22, 2005. This device is used for treatment. Initial product history search conducted on september 28th 2016 revealed no additional complaints against the related part and lot combination. Wear and tear has been determined to be the root cause for the instrument fracture.
Patient Sequence No: 1, Text Type: N, H10


[63312871] It is reported that the device was found to be fractured during kit inspection.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001822565-2016-04753
MDR Report Key6190321
Date Received2016-12-20
Date of Report2016-12-20
Date of Event2016-07-28
Date Mfgr Received2016-11-22
Device Manufacturer Date2005-09-22
Date Added to Maude2016-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DR.
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER, INC.
Manufacturer Street1800 WEST CENTER STREET
Manufacturer CityWARSAW IN 46580
Manufacturer CountryUS
Manufacturer Postal Code46580
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameRUSH AWL 2.5MM DIAMETER REAMER
Generic NameELBOW INSTRUMENT
Product CodeHWJ
Date Received2016-12-20
Returned To Mfg2016-08-12
Model NumberNA
Catalog Number31810616700
Lot Number60351086
ID NumberNA
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER, INC.
Manufacturer Address1800 WEST CENTER STREET WARSAW IN 46580 US 46580


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.