GOLD-CUT METZ SCISSORS FINE 18 CM AD 561-18-01 / 101261

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-22 for GOLD-CUT METZ SCISSORS FINE 18 CM AD 561-18-01 / 101261 manufactured by Geomed Medizin-technik Gmbh & Co. Kg.

Event Text Entries

[63489591] The device history record, including acceptance testing results, was reviewed and found to conform to specifications. The device was subject to microscopic inspection and no signs of excessive force or pressure were found. Due to all of the above, in addition to the fact that the device was returned incomplete, i. E. , the broken-off tip is not available for inspection, it has not been possible to establish the cause of failure.
Patient Sequence No: 1, Text Type: N, H10


[63489592] According to the dealer, the instrument tip was broken during orthopedic surgery. No broken parts inside the patient. No harm done.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010168-2016-00003
MDR Report Key6197784
Date Received2016-12-22
Date of Report2016-12-22
Date Mfgr Received2016-11-22
Device Manufacturer Date2013-11-08
Date Added to Maude2016-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactHANNO HAUG
Manufacturer StreetLUDWIGSTALER STRASSE 27
Manufacturer CityTUTTLINGEN, BADEN WUERTTEMBERG 78532
Manufacturer CountryGM
Manufacturer Postal78532
Manufacturer Phone746193550
Manufacturer G1GEOMED MEDIZIN-TECHNIK GMBH & CO. KG
Manufacturer StreetLUDWIGSTALER STRASSE 27
Manufacturer CityTUTTLINGEN, BADEN-WUERTTEMBERG 78532
Manufacturer CountryGM
Manufacturer Postal Code78532
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameGOLD-CUT METZ SCISSORS FINE 18 CM
Generic NameSURGICAL SCISSORS
Product CodeHRR
Date Received2016-12-22
Returned To Mfg2016-11-24
Catalog NumberAD 561-18-01 / 101261
Lot Number136
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGEOMED MEDIZIN-TECHNIK GMBH & CO. KG
Manufacturer AddressLUDWIGSTALER STRASSE 27 TUTTLINGEN, BADEN-WUERTTEMBERG 78532 GM 78532


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.