LEICA M525 F50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-12-22 for LEICA M525 F50 manufactured by Leica Microsystems (schweiz) Ag.

Event Text Entries

[63073429] An investigation of the incident is currently underway and a follow up will be submitted should additional information become available following the investigation.
Patient Sequence No: 1, Text Type: N, H10


[63073430] Leica microsystems (b)(4) received a complaint on (b)(6) 2016 from (b)(6) stating that in the middle of a surgical procedure, both main and back-up illumination lamps of the leica m525 f50 surgical microscope did not work. The surgeon had to stop the surgery and called the biomedical staff for assistance. The biomedical staff confirmed that both xenon lamp ignitors malfunctioned and the surgical microscope was repaired. The surgical microscope was used again to complete the surgery.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003974370-2016-00013
MDR Report Key6197931
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-12-22
Date of Report2016-12-13
Date of Event2016-12-12
Date Mfgr Received2016-12-13
Device Manufacturer Date2015-07-16
Date Added to Maude2016-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROLAND JEHLE
Manufacturer StreetMAX SCHMIDHEINY STRASSE 201
Manufacturer CityHEERBRUGG, ST. GALLEN 9435
Manufacturer CountrySZ
Manufacturer Postal9435
Manufacturer Phone717263216
Manufacturer G1LEICA INSTRUMENTS (SINGAPORE) PTE LTD
Manufacturer Street12 TEBAN GARDENS CRESCENT
Manufacturer CitySINGAPORE, 608924
Manufacturer CountrySN
Manufacturer Postal Code608924
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA M525 F50
Generic NameSURGICAL MICROSCOPE
Product CodeEPT
Date Received2016-12-22
Model NumberM525 F50
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA MICROSYSTEMS (SCHWEIZ) AG
Manufacturer AddressMAX SCHMIDHEINY STRASSE 201 HEERBRUGG, ST. GALLEN 9435 SZ 9435


Patients

Patient NumberTreatmentOutcomeDate
10 2016-12-22

© 2026 FDA.report