DRIVE DEVILVISS WHEELCHAIR BLS24FBD-SF

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-12-20 for DRIVE DEVILVISS WHEELCHAIR BLS24FBD-SF manufactured by Drive Devilbiss Healthcare / Medical Depot.

Event Text Entries

[63212663] Patient placed hands around seat frame of wheelchair to steady himself as he sat down in the wheelchair. When his weight came down on the seat, the seat dropped approximately 1/2 to 1 inch pinching his fingers under the wheelchair seat. He automatically pulled his hand from the trapped position when the fifth digit, on the palmar side of hand, was lacerated on an exposed screw. This screw was used to secure the seat fabric to the seat frame. The seat is on a sliding mechanism which allows the wheelchair to be folded for storage. When the chair was opened, the sliding mechanism did not slide to the full down position which allowed the wheelchair seat to drop when the patient sat on the seat.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5066838
MDR Report Key6198011
Date Received2016-12-20
Date of Report2016-12-20
Date of Event2016-12-01
Date Added to Maude2016-12-22
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDRIVE DEVILVISS WHEELCHAIR
Generic NameWHEELCHAIR
Product CodeIMP
Date Received2016-12-20
Model NumberBLS24FBD-SF
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDRIVE DEVILBISS HEALTHCARE / MEDICAL DEPOT
Manufacturer AddressPORT WASHINGTON NY 11050 US 11050


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-12-20

© 2026 FDA.report