HUDSON TRACH ADAPTOR 2415-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-01-06 for HUDSON TRACH ADAPTOR 2415-01 manufactured by Teleflex Medical.

Event Text Entries

[64065965] (b)(4). The investigation into this complaint is still in progress.
Patient Sequence No: 1, Text Type: N, H10


[64065966] Customer complaint alleges "tubing on the cannula has holes. " alleged defect detected during use. No harm or injury to the patient. Patient condition reported as "fine. "
Patient Sequence No: 1, Text Type: D, B5


[69118560] (b)(4). The actual sample involved in the reported complaint was not returned for evaluation; therefore, the complaint could not be confirmed. In the current manufacturing procedure 100% leak testing is conducted at the assembly area; therefore, any defects would be detected prior to release from the manufacturing facility. As the actual sample was not returned, ten samples of the same catalog number were selected from current production at the manufacturing facility. A visual exam was performed and no defects were observed. In addition, the samples were leak tested and no issues were detected. It is possible that this defect occurred due to mishandling of the product. The failure was reported during use on a patient. The tube must be soft and flexible in order to meet the hfnc oxygen therapy and user requirement; thus it is prone to tears and pin holes when it is over pulled and stretched. The instructions for use (ifu) mentions that the tubing should not be stretched to remove condensation, or damage/malfunction may occur.
Patient Sequence No: 1, Text Type: N, H10


[69118561] Customer complaint alleges "tubing on the cannula has holes. " alleged defect detected during use. No harm or injury to the patient. Patient condition reported as "fine".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8040412-2017-00019
MDR Report Key6229380
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-01-06
Date of Report2016-12-22
Date of Event2016-12-20
Date Mfgr Received2017-02-16
Date Added to Maude2017-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPO BOX 28, KAMUNTING INDUSTRIAL ESTATE
Manufacturer CityPERAK, WEST MALAYSIA 34600
Manufacturer CountryMY
Manufacturer Postal Code34600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON TRACH ADAPTOR
Generic NameCONNECTOR, AIRWAY (EXTENSION)
Product CodeBZA
Date Received2017-01-06
Catalog Number2415-01
Lot NumberUNKNOWN
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressPERAK, WEST MALAYSIA


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.