ETEST? VORICONAZOLE 412490

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-01-06 for ETEST? VORICONAZOLE 412490 manufactured by Biomerieux, Sa.

Event Text Entries

[64132872] A customer in (b)(6) reported a discrepant result (out-of-range-low) for candida albicans, atcc? 90028, associated with etest? Voriconazole vo 32 ww s30. The customer reported obtaining an unusual shape result, and the ellipse meets the strip under the lower mic. There are three (3) patients involved in this event with three (3) different sample types; pus, trachea aspiration and catheter. The customer reported the iqc result is not good; the expected range is 0. 004 - 0. 016 and a mic result obtain was lower than 0. 002. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There may be a potential for adverse event if the event were to reoccur while testing a patient isolate; therefore this event is being reported as a malfunction. Culture submittal was requested by biom? Rieux for internal investigation. Biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


[75934259] Biom? Rieux conducted an internal investigation: biom? Rieux performed an investigation on customer strips of lot 1004803990 and the retained samples from lot 1004803990 and lot 1005016560 mentioned by customer in parallel with another batch as a reference. We did not reproduce the issue observed by the customer on atcc strain 90028. The conformity of batches 1004803990 and 1005016560 mentioned by customer was be confirmed with two (2) others qc stains whose candida parapsilosis atcc 22019 and candida krusei atcc 6258. On three (3) clinical strains, we reproduced the customers results with mic very low or ellipse which ends below the strip (less than 0. 002) randomly. The strains were susceptible. We confirmed the variability of results obtained by the customer. But the result is considered as acceptable. These results are linked to normal variability intra or inter-batches for antifungals. Indeed, this variability is accepted within two (2) doubling dilutions ((b)(4)) unlike antibiotics. However, these strains have mic very low at the limit of etest? Strip mic. These strains are very susceptible. In this case the real mic do not have a clinical interest. It's the same thing with high level of mic. The variability increase at the limit of the strips (close to 0. 002 or close to 32mg/l).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2017-00003
MDR Report Key6230904
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-01-06
Date of Report2017-05-10
Date Mfgr Received2017-04-14
Device Manufacturer Date2016-04-15
Date Added to Maude2017-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE, 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? VORICONAZOLE
Generic NameETEST? VORICONAZOLE
Product CodeJWY
Date Received2017-01-06
Catalog Number412490
Lot Number1004803990
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE, 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.