HEMOCHRON DIRECTCHECK ACT-LR ABNORMAL DCJLR-A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-01-11 for HEMOCHRON DIRECTCHECK ACT-LR ABNORMAL DCJLR-A manufactured by Accriva Diagnostics.

Event Text Entries

[64557485] (b)(4). Accriva diagnostics has requested all data required for form 3500a.
Patient Sequence No: 1, Text Type: N, H10


[64557486] Healthcare professional reported that an end user sustained an injury during reconstitution of a directcheck quality control. This control is packaged in a glass ampule inside a crushable plastic dropper vial containing diluent. The end-user was wearing gloves but did not utilize the protective sleeve provided with the product. The purpose of the sleeve is to safeguard the end user against potential injury during reconstitution of the control. The end user sustained a small cut to his right thumb, which was caused by a glass shard protruding through the dropper vial. The end user washed the affected area with soap and water, cleaned the cut with an alcohol wipe and applied a band aid. No further medical care was reportedly provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002721930-2017-00001
MDR Report Key6242792
Report SourceHEALTH PROFESSIONAL
Date Received2017-01-11
Date of Report2017-01-09
Date of Event2017-01-09
Date Mfgr Received2017-01-09
Device Manufacturer Date2016-04-12
Date Added to Maude2017-01-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJON MCDERMED
Manufacturer Street6260 SEQUENCE DRIVE
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582632490
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEMOCHRON DIRECTCHECK ACT-LR ABNORMAL
Generic NamePLASMA, COAGULATION CONTROL
Product CodeGGN
Date Received2017-01-11
Model NumberDCJLR-A
Catalog NumberDCJLR-A
Lot NumberD6DLA020
Device Expiration Date2017-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerACCRIVA DIAGNOSTICS
Manufacturer Address6260 SEQUENCE DRIVE SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.