TRIAGE DOA 25 TEST 92590

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-01-11 for TRIAGE DOA 25 TEST 92590 manufactured by Alere San Diego, Inc..

Event Text Entries

[65038893] Investigation conclusion: investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[65038894] Report received of conflicting results provided by triage doa 25 test. Event occurred in (b)(6). Customer alleges a conflicting result has already been seen in more than one case and the issue was observed not at a specific analyte (i. E. , depending on cases, conflicting results have been seen in different analytes). This customer's institute is a hospital particularly focusing on clinical trials and has been using this product for a long time. When using this particular lot (#382197), the customer has been witnessing inconsistent width of lines (narrow / wide lines), as well as conflicting test results (positive and negative) for identical samples. No reported adverse patient sequela. No additional information provided.
Patient Sequence No: 1, Text Type: D, B5


[65930842] Investigation/conclusion: a review was completed of the batch record for lot 382197. Lot met all final release specifications. In-house testing of 29 drug free urine donors on the complaint lot 382197 did not replicate the complaint. All devices tested provided expected negative results. No product deficiency was established.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2017-00014
MDR Report Key6244283
Date Received2017-01-11
Date of Report2016-12-13
Date of Event2016-11-17
Date Mfgr Received2017-01-16
Date Added to Maude2017-01-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE DOA 25 TEST
Generic NameDRUG OF ABUSE TEST
Product CodeMGX
Date Received2017-01-11
Model Number92590
Lot Number382197
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-11