16MM X 100MM X 8ML BD VACUTAINER? PLUS PLASTIC COLLECTION KIT 364946

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2017-01-12 for 16MM X 100MM X 8ML BD VACUTAINER? PLUS PLASTIC COLLECTION KIT 364946 manufactured by Becton, Dickinson & Co..

Event Text Entries

[64668476] Medical device brand name: 16 mm x 100 mm x 8 ml bd vacutainer? Plus plastic collection kit for ua chemistry urine testing. With evacuated conical bottom tube, urinalysis preservative and cup. Sterile. (b)(4). Device evaluation: result - one used sample was returned for evaluation. The cup was inspected for defects with no visible defects identified. The cup was then filled with water, and closed, shaken vigorously in an attempt to produce leaks, with no leaking occurring. A review of the device history record cannot be completed as the lot number was not provided for this incident. Conclusion - the complaint for urine leakage was not confirmed. Without a sample, an absolute root cause for this incident cannot be determined.
Patient Sequence No: 1, Text Type: N, H10


[64668477] It was reported that the clinician was accessing the bd closed urine collection cup from the suspect device and urine splashed into his/her eye.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1917413-2017-00001
MDR Report Key6246995
Report SourceOTHER,USER FACILITY
Date Received2017-01-12
Date of Report2017-01-05
Date of Event2016-12-20
Date Mfgr Received2016-12-20
Date Added to Maude2017-01-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652845
Manufacturer G1BECTON, DICKINSON & CO.
Manufacturer Street150 SOUTH 1ST AVENUE
Manufacturer CityBROKEN BOW NE 68822
Manufacturer CountryUS
Manufacturer Postal Code68822
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name16MM X 100MM X 8ML BD VACUTAINER? PLUS PLASTIC COLLECTION KIT
Generic NameCOLLECTION KIT
Product CodeJXA
Date Received2017-01-12
Returned To Mfg2017-01-04
Catalog Number364946
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerBECTON, DICKINSON & CO.
Manufacturer Address150 SOUTH 1ST AVENUE BROKEN BOW NE 68822 US 68822


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-01-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.