MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-01-16 for UNKNOWN_INSTRUMENTSTIRE_PRODUCT UNK_ICO manufactured by Stryker Instruments-kalamazoo.
[65347946]
A follow up report will be filed once the quality investigation is complete.
Patient Sequence No: 1, Text Type: N, H10
[65347947]
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment. It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
[67450207]
Disassembly and visual inspection of the associated device (b)(4) noted the driveshaft assembly was damaged. This may have potentially caused the reported event as damage may limit the design function of the device. The affected driveshaft assembly was replaced and the drill repaired. As the bur could not be identified and that only a partial piece of a bur was returned, further evaluation of this partial piece of bur was not possible.
Patient Sequence No: 1, Text Type: N, H10
[67450208]
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment. It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0001811755-2017-00062 |
MDR Report Key | 6253169 |
Date Received | 2017-01-16 |
Date of Report | 2017-02-15 |
Date of Event | 2016-12-20 |
Date Mfgr Received | 2017-02-03 |
Date Added to Maude | 2017-01-16 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. UNA BARRY |
Manufacturer Street | INSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK |
Manufacturer City | CARRIGTWOHILL NA |
Manufacturer Postal | NA |
Manufacturer Phone | 214532900 |
Manufacturer G1 | STRYKER INSTRUMENTS-IRELAND |
Manufacturer Street | INSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK |
Manufacturer City | CARRIGTWOHILL NA |
Manufacturer Postal Code | NA |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | UNKNOWN_INSTRUMENTSTIRE_PRODUCT |
Generic Name | UNKNOWN |
Product Code | DZJ |
Date Received | 2017-01-16 |
Returned To Mfg | 2016-12-19 |
Catalog Number | UNK_ICO |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STRYKER INSTRUMENTS-KALAMAZOO |
Manufacturer Address | 4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-01-16 |