TM-400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-01-18 for TM-400 manufactured by Ito Co., Ltd. Tsukuba Factory.

Event Text Entries

[65565855] The subject device was discarded at the event site. Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[65565856] During a procedure, the traction rope of the subject tm-400 device got stuck. Even after waiting a while to try to pull out the rope, the rope was still stuck and could not be pulled out. After turning off the power and trying to resume the procedure, the rope could be reeled up and a smooth traction was resumed. When the traction power level was set from 0 to 60%, the rope got stuck instantaneously and then started traction with a high speed (a speed of 60 to 100% traction power). When the 90kg setting was selected, the rope became stuck without exception.
Patient Sequence No: 1, Text Type: D, B5


[73684965] Ito examined the device history record for the subject tm-400 device. There were no problems observed during the manufacturing or testing noted in the dhr. The user facility did not disclose the patient's information.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9614750-2017-00003
MDR Report Key6261009
Date Received2017-01-18
Date of Report2017-01-18
Date of Event2016-03-14
Date Mfgr Received2016-03-14
Device Manufacturer Date2015-01-23
Date Added to Maude2017-01-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KENNETH BLOCK
Manufacturer Street1201 RICHARDSON DR. SUITE 160
Manufacturer CityRICHARDSON TX 75080
Manufacturer CountryUS
Manufacturer Postal75080
Manufacturer Phone9724809554
Manufacturer G1ITO CO., LTD. TSUKUBA FACTORY
Manufacturer Street3-39-14 UZURANO AMI-MACHI
Manufacturer CityINASHIKI-GUN, IBARAKI 300-0341
Manufacturer CountryJA
Manufacturer Postal Code300-0341
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTM-400
Generic NameEQUIPMENT, TRACTION, POWERED
Product CodeITH
Date Received2017-01-18
Model NumberTM-400
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerITO CO., LTD. TSUKUBA FACTORY
Manufacturer Address3-39-14 UZURANO AMI-MACHI INASHIKI-GUN, IBARAKI 300-0341 JA 300-0341


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-01-18