BUSH URETERAL ILLUMINATING CATHETER SET

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-01-19 for BUSH URETERAL ILLUMINATING CATHETER SET manufactured by Cook Incorporated.

Event Text Entries

[65245254]
Patient Sequence No: 1, Text Type: N, H10


[65245255] During a surgical procedure, a ureteral laminating stent was plugged into light source and after 15 to 20 seconds after being plugged in, staff noted smoke coming out of the light source hole above where the stent was plugged into. Surgeon stated unable to see stent lighted and surgeon started to unplug it. After turning off, smoke came out of both holes where it was unplugged. Staff person reported visually saw bright red center at end of stent where it was plugged into light source. Biomed noted event as "unit smoked when user connected ureteral illuminating catheter set. Cook medical; lot #6998468; ref # (b)(4). Tech was able to duplicate the problem when he connected the catheter to the light source. Unit worked ok with other head set. Contacted manufacturer and reported the problem and a representative picked up the catheter for an investigation. In follow up to the investigation, cook reported that there was no indication that a design or process related failure mode contributed to the event a definitive root cause could not be established.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6263152
MDR Report Key6263152
Date Received2017-01-19
Date of Report2017-01-18
Date of Event2016-11-11
Report Date2017-01-18
Date Reported to FDA2017-01-18
Date Reported to Mfgr2017-01-18
Date Added to Maude2017-01-19
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBUSH URETERAL ILLUMINATING CATHETER SET
Generic NameLIGHT, CATHETER, FIBEROPTIC, GLASS, URETERAL
Product CodeFCS
Date Received2017-01-19
Returned To Mfg2016-11-22
Lot Number6998468
OperatorNURSE
Device AvailabilityR
Device Age1 DY
Device Eval'ed by Mfgr*
Device Sequence No0
Device Event Key0
ManufacturerCOOK INCORPORATED
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-19

© 2026 FDA.report