SCHUHKNECHT SUC. TUBE 24GA 166/355/024

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-01-20 for SCHUHKNECHT SUC. TUBE 24GA 166/355/024 manufactured by Erchinger Medtechnology Gmbh & Co.kg.

Event Text Entries

[65782701] The suction tubes have not been returned completely. The front thin tube is missing, which has broken off. We have examined the suction tubes with the following result: the inner tube is broken before the soldering point on the front thin tube of the inner tube. The tube is not broken at the solder joint itself. The inner tube is bent so that the product has been leveraged. This is obvious, so that only the inner tube has bent. The outer tube is undamaged. The production documents (dhr) are available in full: drawing number: (b)(4) rev. A. Inspection plan (b)(4) dated from june 24, 2015. Route card (b)(4) dated from april 09, 2015 with part list. The examination of the production documents showed the correct production of the products.
Patient Sequence No: 1, Text Type: N, H10


[65782702] Customer initially reports schuhknecht suction tube 24ga broke at hub during use in surgery. (b)(6) 2016 customer reports device broke while suctioning the nasal cavity. No harm done, no further information available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807421-2016-00001
MDR Report Key6266061
Date Received2017-01-20
Date Mfgr Received2016-09-21
Date Added to Maude2017-01-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMO RACK
Manufacturer StreetRUDOLF-DIESEL-STRASSE 20
Manufacturer CityTUTTLINGEN, BADEN-W 78532
Manufacturer CountryGM
Manufacturer Postal78532
Manufacturer Phone461966250
Manufacturer G1ERCHINGER MEDTECHNOLOGY GMBH & CO.KG
Manufacturer StreetRUDOLF-DIESEL-STRASSE 20
Manufacturer CityTUTTLINGEN, BADEN-W 78532
Manufacturer CountryGM
Manufacturer Postal Code78532
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSCHUHKNECHT SUC. TUBE 24GA
Generic NameJARIT
Product CodeJZF
Date Received2017-01-20
Returned To Mfg2016-09-15
Model Number166/355/024
Catalog Number166/355/024
Lot Number20510506
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerERCHINGER MEDTECHNOLOGY GMBH & CO.KG
Manufacturer AddressRUDOLF-DIESEL-STRASSE 20 TUTTLINGEN 78532 US 78532


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-20

© 2024 FDA.report