BN PROSPEC OVLI11 SMN10461865

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-01-20 for BN PROSPEC OVLI11 SMN10461865 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[65469222] The cause of the discordant depressed immunoglobulin g, immunoglobulin a, and immunoglobulin m (igg, iga and igm) results is unknown. The pattern of low results on multiple analytes on the same sample is suggestive of a pre-analytical sample issue. The instrument is performing according to specifications. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


[65469223] Discordant depressed immunoglobulin g, immunoglobulin a, and immunoglobulin m (igg, iga and igm) results were obtained on a patient sample on the bn prospec instrument. The results were reported to the physician. The same sample was later repeated on the same instrument and higher results were obtained and reported. The initial depressed results were regarded as discordant. There is no indication that patient treatment was altered or prescribed on the basis of the discordant depressed igg, iga and igm results. There was no report of adverse health consequences as a result of the discordant depressed igg, iga and igm results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00006
MDR Report Key6266559
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-01-20
Date of Report2017-01-20
Date of Event2016-12-21
Date Mfgr Received2016-12-22
Device Manufacturer Date2015-08-01
Date Added to Maude2017-01-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, D-35041
Manufacturer CountryGM
Manufacturer Postal CodeD-35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBN PROSPEC
Generic NameBN PROSPEC NEPHELOMETER/PC/SOFTWARE KIT
Product CodeJZW
Date Received2017-01-20
Catalog NumberOVLI11 SMN10461865
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-20