REICHERT/CAMBRIDGE 14364 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2005-08-01 for REICHERT/CAMBRIDGE 14364 * manufactured by Reichert Opthalmic Instruments.

Event Text Entries

[385662] The patient was seated upright in the ophthalmology clinic examination chair. The slit lamp microscope table was placed over the patient's legs by the physician in preparation for an eye examination. The physician pushed the button on the ophthalmic console/instrument stand in order to raise the chair to the level needed for the exam. The chair began to rise. The chair continued to rise despite the physician having released the button. The patient's thigh was pinned between the chair and the slit lamp table. A technician in the room unplugged the electrical cord that was connected to the chair via a relay box from the console/stand. The chair stopped rising. The technician then plugged the cord back in, the physician pushed the button for the chair to lower. The chair lowered, releasing the patient's leg.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number627323
MDR Report Key627323
Date Received2005-08-01
Date of Report2005-08-01
Date of Event2005-07-13
Report Date2005-08-01
Date Reported to FDA2005-08-01
Date Added to Maude2005-08-17
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameREICHERT/CAMBRIDGE
Generic NameOPHTHALMIC CONSOLE
Date Received2005-08-01
Model Number14364
Catalog Number*
Lot Number*
ID Number*
OperatorPHYSICIAN
Device AvailabilityY
Device Age13 YR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key616949
ManufacturerREICHERT OPTHALMIC INSTRUMENTS
Manufacturer Address3374 WALDEN AVENUE DEPEW NY 14043 US

Device Sequence Number: 2

Brand NameREICHERT/CAMBRIDGE
Generic NameOPHTHALMIC INSTRUMENT STAND
Product CodeHMF
Date Received2005-08-01
Model Number14363
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age13 YRS
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key616950
ManufacturerREICHERT OPHTHALMIC INSTRUMENTS
Manufacturer Address3374 WALDEN AVE DEPEW NY 14043 US

Device Sequence Number: 3

Brand NameREICHERT/CAMBRIDGE
Generic NameOPHTHALMIC CHAIR
Product CodeHME
Date Received2005-08-01
Model Number14360
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age13 YR
Implant FlagN
Date Removed*
Device Sequence No3
Device Event Key616951
ManufacturerREICHERT OPHTHALMIC INSTRUMENTS
Manufacturer Address3374 WALDEN AVE DEPEW NY 14043 US


Patients

Patient NumberTreatmentOutcomeDate
10 2005-08-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.