DIMENSION? RF420 SMN 10444895

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-01-25 for DIMENSION? RF420 SMN 10444895 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[65736336] Analysis of the information provided indicate that the cause for the discordant elevated mmb result is unknown. The elevated mmb values were repeatable and discordant with other cardiac test values. The instructions for use for the mass creatine kinase mb isoenzyme flex reagent cartridge contains the following information: "patient samples may contain heterophilic antibodies that could react in immunoassays to give falsely elevated or depressed results. This assay has been designed to minimize interference from heterophilic antibodies. Nevertheless, complete elimination of this interference cannot be guaranteed. A test result that is inconsistent with the clinical picture and patient history should be interpreted with caution. " the device is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[65736337] Discordant, elevated mass creatine kinase (mmb) results were obtained on a patient's samples on the dimension exl system. The results were reported to the physician who questioned the results. The patient's samples were retested at an alternate facility with a non-siemens alternate methodology and lower results were obtained. Patient treatment was not altered or prescribed on the basis of the discordant elevated mass creatine kinase (mmb) results. There are no reports of patient intervention or adverse health consequences due to the discordant elevated mass creatine kinase (mmb) results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00028
MDR Report Key6277344
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-01-25
Date of Report2017-01-25
Date of Event2016-12-25
Date Mfgr Received2016-12-30
Device Manufacturer Date2016-06-15
Date Added to Maude2017-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal Code197146101
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION?
Generic NameDIMENSION? MMB MASS CREATINE KINASE MB ISOENZYME FLEX? REAGENT CARTRIDGE
Product CodeJHY
Date Received2017-01-25
Catalog NumberRF420 SMN 10444895
Lot NumberEB7166
Device Expiration Date2017-06-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.