Event Text Entries

[65772019] A steris service technician arrived onsite to inspect the washer. The technician inspected the washer and observed the drifting condition of the unload washer door as reported by the user facility. The technician stated that the drifting of the door closing was slow which should have allowed time for the employee to remove their arm before the door contacted it. The technician further inspected the washer and found that the hinge pins required adjustment. The technician adjusted the hinge pins and returned the washer to service. The load and unload doors on the 1227 cart washer are manual hinged doors. Since the doors are manual hinged doors, the operator of the device controls the speed of the door closure. The operator manual states (pp. 1-1), "keep hands and fingers away from door or any moving parts to prevent crushing. " a review of 1227 cart washer complaints was completed and revealed the event to be an isolated occurrence.
Patient Sequence No: 1, Text Type: N, H10


[65772020] The user facility reported that the unload door on their 1227 cart washer drifted and closed on an employee's arm. The employee sought and received medical treatment at the facility's emergency department. The user facility stated that the employee is back to work.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680353-2017-00010
MDR Report Key6277702
Date Received2017-01-25
Date of Report2017-01-25
Date of Event2016-12-28
Date Mfgr Received2016-12-28
Date Added to Maude2017-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC 8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameRELIANCE 1227 CART WASHER
Generic NameWASHER
Product CodeFLH
Date Received2017-01-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-01-25