TRIAGE PROFILER SOB PANEL 97300EU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional, report with the FDA on 2017-01-25 for TRIAGE PROFILER SOB PANEL 97300EU manufactured by Alere San Diego, Inc..

Event Text Entries

[65773405] Customer's complaint was not replicated with in-house testing of retain lot w62214. No issues with d-dimer recovery were observed. Manufacturing batch records for lot w62214 were reviewed and found that the lot met release specifications. Based on the information available, there is no indication of a product deficiency and no corrective action is required. Although this catalog number (97300eu) is not approved in the united states, this event is being reported as the device is same/similar to catalog number 97300, 510(k) number k080269.
Patient Sequence No: 1, Text Type: N, H10


[65773406] The customer porvided the following events occurring on (b)(6) 2017. Patient presented with mid thoracic back pain that was worse on inspiration. No calf tenderness or swelling. Occurred three days after air travel. Triage d-dimer result of less than 100 ng/ml. Facility unable to provide triage d-dimer cutoff but customer stated this was a normal result. Four and a half hours later, siemens innovance d-dimer result of 950 ng/ml (1. 9mg/l feu = 950 ng/ml ddu). Facility siemens d-dimer cutoff is less than 0. 5 mg/l. No information on treatment or intervention available. No additional information available.
Patient Sequence No: 1, Text Type: D, B5


[75235795] Corrections: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2017-00023
MDR Report Key6278084
Report SourceCONSUMER,HEALTH PROFESSIONAL,
Date Received2017-01-25
Date of Report2017-01-17
Date of Event2017-01-17
Date Mfgr Received2017-01-17
Date Added to Maude2017-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE PROFILER SOB PANEL
Generic NameCARDIOPROFILER PANEL
Product CodeDAP
Date Received2017-01-25
Catalog Number97300EU
Lot NumberW62214
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2017-01-25