SAFEBASICS SALIVA EJECTOR 700

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-01-26 for SAFEBASICS SALIVA EJECTOR 700 manufactured by Amd Medicom Inc..

Event Text Entries

[66076940] A complaint was reported to amd medicom inc. On 05-dec-2016 for medicom brand product saliva ejector item code# 700 lot number #100. 16. The mask is classified as a class 1 device, under fda code dyn, whose intended use is to supply power to and serve as a base for other dental devices, such as a dental hand piece, a dental operating light, and an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to dental chair. The following information was initially reported: "saliva ejector tip fell off possibly in patient and tips are loose and not staying on. First time this happened. Client has been using the product for years" on 07-dec-2016 client was contacted by amd medicom by phone for further information. (b)(6) from the (b)(6) hospital anesthesia department mentioned that the patient was undergoing an endoscopy procedure when the incident happened. They are still unsure if the tip fell off into the patient but they could not locate the tip anywhere and he believes there is a chance that the tip was swallowed by the patient. He mentioned the patient would have symptoms if they aspirated the tip into the lungs. He confirmed that the patient had no symptoms. He mentioned that the tips are normally stored in the bags on a shelf in the hospital. The complaint received by amd medicom has been logged as a product quality complaint under (b)(4). A request for investigation was made to the manufacturing facility, through capa (b)(4). Investigation: - review of amd medicom quality complaint records confirmed that no similar issues had been reported for item code # 700, under lot # 100. 16. - retain samples were inspected by amd medicom. The blue cap was not easily removable. - return samples were received from end-user related to the nature of the complaint. Sixty (60) units (i. E. Saliva ejectors) were received and inspected, only 09 units had blue cap that were potentially removable when manipulated. - return samples were sent to the manufacturing facility for further evaluation. Root cause: - the root cause was identified to be related to line clearance. Some removable tips from the previous production were not removed from the production line. Corrective actions: - during every production change from the manufacture of removable to fixed tips, first two minutes of production needs to be removed from the production lot. -the manufacturing facility procedure mdq/8. 2. 4/a "products monitoring and measurement" has been updated to reflect this change. In conclusion, the risk associated to the nature of complaint is considered low if it is a short term exposure to pvc which is the material the saliva ejectors are made of. Furthermore, no other complaints of this nature have been reported for this product and lot number.
Patient Sequence No: 1, Text Type: N, H10


[66076941] Saliva ejector tip fell off possibly in patient and tips are loose and not staying on. First time this happened. Been using the product for years. Noted several in this package that tip easily removed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611959-2017-00001
MDR Report Key6281928
Date Received2017-01-26
Date of Report2017-01-26
Date of Event2016-12-05
Date Mfgr Received2016-12-05
Date Added to Maude2017-01-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DOMINIQUE ABECASSIS
Manufacturer Street1200 55 TH AVENUE
Manufacturer CityMONTREAL, QUEBEC H8T 3J8
Manufacturer CountryCA
Manufacturer PostalH8T 3J8
Manufacturer Phone6366262
Manufacturer G1WORLD WORK SRL
Manufacturer StreetVIA DEL PROGRESSO 47 MONTEBELLO
Manufacturer CityVICENZA, 36054
Manufacturer CountryIT
Manufacturer Postal Code36054
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSAFEBASICS SALIVA EJECTOR
Generic NameSALIVA EJECTOR
Product CodeDYN
Date Received2017-01-26
Returned To Mfg2016-12-15
Catalog Number700
Lot Number100.16
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAMD MEDICOM INC.
Manufacturer Address1200 55TH AVENUE MONTREAL, QUEBEC H8T 3J8 CA H8T 3J8


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-01-26

© 2026 FDA.report