CRYOVALVE SG AORTIC VALVE & CONDUIT SGAV00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2017-02-02 for CRYOVALVE SG AORTIC VALVE & CONDUIT SGAV00 manufactured by Cryolife, Inc. ? Kennesaw.

Event Text Entries

[66336873] This investigation is currently ongoing. Any additional information will be provided in the follow-up report.
Patient Sequence No: 1, Text Type: N, H10


[66336874] According to the report from the cryolife representative, "[the surgeon] asked me to come to his office to discuss the allograft sizing issues over the last 6 months. He has observed, and has documented, complete with angiograms of patients, that some of the smaller size av00's are becoming stenotic post operatively. In addition, he has observed that the smaller size av00's also appear to be shrinking postoperatively. He noted one incident where an implanted 9mm av00 was measured to be a 6mm postoperatively. I asked if someone from cryolife could contact him if they had more questions regarding these observations, and [the surgeon] said 'oh sure. ' [the surgeon] is concerned with the sizing issue that has occurred several times in the last 6 months.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1063481-2017-00002
MDR Report Key6295982
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2017-02-02
Date of Report2018-03-09
Date of Event2016-12-22
Date Facility Aware2017-01-03
Date Mfgr Received2017-01-03
Date Added to Maude2017-02-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROCHELLE MANEY
Manufacturer Street1655 ROBERTS BLVD., NW
Manufacturer CityKENNESAW GA 30144
Manufacturer CountryUS
Manufacturer Postal30144
Manufacturer G1CRYOLIFE, INC.
Manufacturer Street1655 ROBERTS BLVD. NW
Manufacturer CityKENNESAW GA 30144
Manufacturer CountryUS
Manufacturer Postal Code30144
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCRYOVALVE SG AORTIC VALVE & CONDUIT
Generic NameHEART VALVE, MORE THAN MINIMALLY MANIPULATED ALLOGRAFT
Product CodeOHA
Date Received2017-02-02
Model NumberSGAV00
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCRYOLIFE, INC. ? KENNESAW
Manufacturer Address1655 ROBERTS BLVD. NW KENNESAW GA 30144 US 30144


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-02-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.