CALIBRATION DRILL V040925000001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-02 for CALIBRATION DRILL V040925000001 manufactured by Dentsply Vdw Gmbh.

Event Text Entries

[66687961] There has been a previous report received where this malfunction in a similar device resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21cfr part 803. Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply (b)(4). The involved drill is actually broken in the middle of the smaller step. No material defect was found during of the analysis of the rupture pattern. Unused available drills have been measured and meet drawing specifications. Nothing unusual to report was found during dhr review. Root causes are not identified. We will track this kind of event and monitor the trend.
Patient Sequence No: 1, Text Type: N, H10


[66687962] In this event it was reported that a calibration drill separated. Multiple unsuccessful attempts were made to obtain the patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611053-2016-00084
MDR Report Key6297233
Date Received2017-02-02
Date of Report2017-02-02
Date Mfgr Received2016-06-30
Date Added to Maude2017-02-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. HELEN LEWIS
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494229
Manufacturer G1DENTSPLY MAILLEFER
Manufacturer StreetCHEMIN DU VERGER 3
Manufacturer CityBALLAIGUES, 1338
Manufacturer CountrySZ
Manufacturer Postal Code1338
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCALIBRATION DRILL
Generic NameDRILL, DENTAL, INTRAORAL
Product CodeDZA
Date Received2017-02-02
Returned To Mfg2016-06-30
Model NumberNA
Catalog NumberV040925000001
Lot Number1150461
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDENTSPLY VDW GMBH
Manufacturer AddressBAYERWALDSTRASSE 15 MUNICH, 81737 GM 81737


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-02