MAUDE MDR 6299125

MDR report key
6299125
Report number
1644019-2017-00111
Event key
0
Event type
3
Date of event
2017-01-18
Date received
2017-02-03
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
2
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MS. RITA LOPEZ
Address
6201 SOUTH FREEWAY MAIL STOP AB2-6 FORT WORTH TX 76134 US
Phone
817-817-8175
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT)TUBING, NONINVASIVEALCON RESEARCH, LTD. - HOUSTONGAZNA80657509571926002HR N

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12017-02-0301. O

Event Narratives#

N

Patient 1

A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

D

Patient 1

A NURSE MANAGER REPORTED THAT DURING A VITRECTOMY PROCEDURE, THE SURGEON INDICATED THAT THE CANNULA DISCONNECTED FROM THE SYRINGE WHEN EXERTING PRESSURE TO INFUSE THE SILICONE OIL. THIS OCCURRED WITH TWO DIFFERENT UNITS. THE PROCEDURE WAS COMPLETED BY INFUSING THE SILICONE OIL SLOWLY AND WITH THE ASSISTANCE OF FORCEPS TO HOLD THE CONNECTION. THE PRODUCT SAMPLES WERE NOT RETAINED. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED; HOWEVER, NONE HAS BEEN RECEIVED TO DATE.

N

Patient 1

THE LOT COMPLAINT HISTORY WAS REVIEWED; THIS IS THE FIRST COMPLAINT FOR THE FINISH GOODS LOT AND FIRST FOR THIS ISSUE. THE DEVICE HISTORY RECORD SHOWED THE PRODUCT WAS RELEASED PER SPECIFICATIONS. THE CUSTOMER DID NOT RETAIN A SAMPLE FOR THIS COMPLAINT REPORT; VISUAL INSPECTION OR FUNCTIONAL TESTING COULD NOT BE CONDUCTED. THE ROOT CAUSE OF THE CUSTOMER'S COMPLAINT COULD NOT BE ESTABLISHED AS A SAMPLE HAS NOT BEEN RECEIVED. POTENTIAL ROOT CAUSES COULD BE RELATED TO SURGICAL TECHNIQUE OR MANUFACTURING RELATED ISSUE, HOWEVER, THIS CANNOT BE CONFIRMED WITH THE INFORMATION AVAILABLE. AFTER A THOROUGH INVESTIGATION OF THIS COMPLAINT, IT HAS BEEN DETERMINED THAT NO ACTION WILL BE TAKEN AT THIS TIME. (B)(4).